Photometric measurement and control system for disinfectant concentrations
- Conforming with standard DPD-method according to AWWA 4500-Cl G/EN ISO 7393-2 for free chlorine
- High accuracy and reproduceability due to automatic zero-value calibration
- Reduced maintenance with cleaning module and high tolerance against fouling
Specifications
Photometric measurement and control system for disinfectant concentrations
Disinfectant measurement | Measuring range [ppm] | Accuracy | |
Ozone | 0.05 | 1.00 | ±0.01 ppm |
HOCL, free chlorine, monochloramine | 0.00 | 1.00 | ±0.01 ppm |
1.00 | 3.00 | ±0.06 ppm | |
3.00 | 5.00 | ±0.2 ppm | |
Chlorine dioxide, iodine, bromine | 0.00 | 2.00 | ±0.02 ppm |
2.00 | 6.00 | ±0.12 ppm |
Response time | 90% of change of excessive Cl in 60 seconds after sample entered flow cell. |
Cycle time free chlorine | 1 … 12 min. |
pH measuring range | 2 … 12 |
Temperature measuring range | -30 … 100 °C |
Transmitter Specifications |
|
Power Supply | 100-240 VAC (±10%), 50/60Hz or 10-36 VDC |
Power consumption | max. 35 VA |
Sample conditions |
|
Water consumption | min. 10 l/h |
Water inlet pressure | 0.15 … 2 bar |
Sample temperature | 5 … 50 °C |
Panel |
|
Dimension | 400 x 850 x 200 mm |
Material | PVC |
Weight | 9.0 kg |
Contact us to place an order.