Continuous measurement of dissolved hydrogen for corrosion monitoring in water steam cycles
- Faraday electrode for automatic or manual verification by electrochemically generated hydrogen concentration
- Self-monitoring of membrane integrity and electrolyte
- Long-life amperometric hydrogen sensor
Specifications
Continuous measurement of dissolved hydrogen for corrosion monitoring in water steam cycles
| Dissolved Hydrogen measurement | |
| Measuring range | 0 to 800 ppb or 0 to 50% saturation |
| Resolution | 0.01 ppb (up to 9.99 ppb) 0.1 ppb (from 10.0 to 99.9 ppb) 1 ppb (from 100 to 800 ppb) |
| Accuracy | +- 5% of reading or +- 0,5 ppb (whichever is greater) |
| Repeatability | +- 1% of reading or +- 0,5 ppb (whichever is greater) |
| Temperature measurement | |
| Measuring range | -30 to 130°C |
| Resolution | 0.1° C |
| Power Supply | |
| Voltage | 100-240 VAC (+- 10%), 50/60 Hz (+- 5%) or 10-36 VDC (+- 10%) |
| Sample conditions | |
| Flow rate | 6 to 20 l/h |
| Temperature | up to 45° C |
| Inlet pressure | up to 1 bar |
| Outlet pressure | pressure free |
| Suspended solids | less than 10 ppm |
| Dimensions | |
| Panel | 280 x 850 x 150 mm |
For further information see datasheet in downloads.