AMI-II Powercon Specific is an advanced online analyser for continuous measurement of specific conductivity in high-purity water systems.
The AMI-II Powercon Specific is a complete monitoring system for the continuous measurement of specific, or total, conductivity in high-purity water. It combines the dedicated AMI-II Powercon transmitter, a high-precision conductivity sensor, flow cell and digital sample flow meter. Selectable temperature compensation supports varying sample qualities and alkalising agents, while automatic zero verification helps maintain measurement accuracy. The analyser provides reliable, automatic and maintenance-free operation for power plants, industrial facilities and water treatment systems.
Specifications
- Selectable temperature compensation for varying sample qualities and alkalising agents.
- Automatic zero verification with an integrated high-precision resistor for accuracy.
- Two-electrode titanium conductivity sensor with a high-precision cell constant and Pt1000 probe.
- Patented slot-lock sensor design for quick and easy installation and release.
- Integrated needle valve and digital flow meter for easy and reliable operation.
- Wet bench-tested at the factory and ready for installation and operation.
AMI-II Powercon Specific Specifications
The AMI-II Powercon Specific includes the AMI-II Powercon transmitter, UP-Con1000-SL conductivity sensor and QV-Flow UP-CON-SL flow cell. The complete system is mounted on a stainless steel panel and is ready for installation and operation.
The transmitter includes two signal outputs for measured values or continuous control, two relays configurable as limit switches or controllers, an alarm relay for summary alarm indication and an input for a potential-free contact.
The flow cell is equipped with a manual flow adjustment valve and digital sample flow meter.
Measuring range
| Conductivity | 0.055 µS/cm to 30 mS/cm |
Sample conditions
| Temperature | 0 to 50 °C |
| Inlet pressure | Maximum 2 bar |
| Outlet pressure | Pressure-free |
Sample connections
| Inlet | Swagelok ¼" tube adapter |
| Outlet | ⅛" adapter for flexible tube, 8 mm diameter |
| Power supply | 100–240 VAC (±10%) or 10–36 VDC |
Contact us to place an order.
Downloads
| Brochure | Product Flyer |
| Manual | Manual |
| Datasheet | Datasheet |