Monitor AMI-II Turbisafe

AMI-II Turbisafe

AMI-II Turbisafe is a cost-efficient online analyser for continuous turbidity measurement in accordance with ISO 7027.

The AMI-II Turbisafe is designed for reliable turbidity monitoring across a broad measuring range of 0.000 to 100 FNU/NTU. Supplied fully tested, calibrated and ready for operation, it features a robust flow cell capable of handling inlet pressures up to 10 bar, helping prevent measurement errors caused by sample degassing.

The system requires no desiccant bags, cuvettes or other consumables, while the flow cell and sensor head can be cleaned quickly for low-maintenance operation. Measurements can be validated or calibrated using a primary formazine standard. Optional features include a second turbidity sensor, sample flow monitoring or control, and an automated cleaning module.

Specifications

  • Continuous turbidity measurement in accordance with ISO 7027.
  • Broad measuring range from 0.000 to 100 FNU/NTU.
  • Withstands inlet pressures up to 10 bar, helping prevent errors caused by sample degassing.
  • No desiccant bags, cuvettes or other consumables required.
  • Flow cell and sensor head can be cleaned quickly for low-maintenance operation.
  • Measurement results can be validated or calibrated using a primary formazine standard.
  • Optional second turbidity sensor, sample flow monitoring or control, and automatic cleaning module.

 

AMI-II Turbisafe Specifications

The AMI-II Turbisafe is a complete online analyser for continuous turbidity measurement in accordance with ISO 7027. It is delivered fully tested, calibrated and ready for installation and operation.

The robust flow cell withstands inlet pressures up to 10 bar, helping prevent measurement errors caused by sample degassing. The system requires no desiccant bags, cuvettes or other consumables, helping minimise ongoing operating costs.

Optional configurations include a second turbidity sensor for monitoring before and after filtration, a sample flow meter or flow controller, and an automatic cleaning module.

 

Turbidity measurement

Measurement principle Nephelometric measurement in accordance with ISO 7027
Measuring range 0.000 to 100 FNU/NTU
Precision ±0.02 FNU (limit of detection)
Calibration standard Primary formazine standard

Measurement accuracy

0 to 1 FNU ±0.01 FNU or ±2%
1 to 10 FNU ±0.1 FNU or ±3%
10 to 100 FNU ±0.5 FNU or ±5%

Sample conditions

Flow rate Approximately 5 to 50 L/h
Sample temperature Up to 40 °C
Inlet pressure Up to 10 bar
Outlet pressure Pressure-free

Power supply

AC version 100–240 VAC (±10%), 50/60 Hz (±5%)

 

Contact us to place an order.

Downloads

Brochure Product Flyer
Manual Manual
Datasheet Datasheet