Automatic, continuous measurement of total chlorine in potable water, sanitary water, cooling water and effluent.
- Continuous online determination of disinfectants based on the DPD colorimetric method (EN ISO 7393-2; APHA 4500-Cl G).
- Measurement values: total chlorine 1, total chlorine 2, dichloramine, flow and if installed pH and temperature.
- Integrated pH measurement with temperature compensation available as option.
Specifications
Automatic, continuous measurement of total chlorine in potable water, sanitary water, cooling water and effluent.
Disinfectant measurement | Measuring range (Standard) [ppm] | Accuracy | |
Total chlorine (tc1 & tc2) | 0.00 | 1.00 | ± 0.01 ppm |
1.00 | 3.00 | ± 0.06 ppm | |
3.00 | 5.00 | ± 0.2 ppm | |
Measuring range (Extended) [ppm] | |||
0.00 | 10 | ± 10 % | |
Dichloramine | Calculated difference of total chlorine 1 and 2. |
Measurement time tc1 | 3 -5 sec. |
Measurement time tc2 | 2 min. |
Cycle time | 3 … 60 min. |
pH measuring range | 2 … 12 |
Temperature measuring range | -30 … 100 °C |
Transmitter Specifications |
|
Power Supply | 100 … 240 VAC (±10%), 50/60Hz or 10-36 VDC |
Power consumption | max. 35 VA |
Sample conditions |
|
Water consumption | min. 10 l/h |
Water inlet pressure | 0.15 … 2 bar |
Sample temperature | 5 … 50 °C |
Panel |
|
Dimension | 400 x 850 x 200 mm |
Material | PVC |
Weight | 9.0 kg |